Design of a Configurable 4-Channel Analog Front-End for EEG Signal Acquisition on 180nm CMOS Process

Duc Hung Le, The-Hung Pham, Cong-Kha Pham

Abstract


In this work, a 4-channel Analog Front-End (AFE) circuit has been proposed for EEG signal recording. For EEG recording systems, the AFE may handle a wide range of sensor inputs with high input impedance, adjustable gain, low noise, and wide bandwidth. The buffer or current-to-voltage converter block (BCV), which can be set to operate as a buffer or a current-to-voltage converter circuit, is positioned between the electrode and the main amplifier stages of the AFE to achieve high input impedance and work with sensor signal types. A chopper capacitively-coupled instrumentation amplifier (CCIA) is positioned after the BCV as the main amplifier stage of the AFE to reduce input-referred noise and balance the impedance of the overall AFE system. A programmable gain amplifier (PGA) is the third stage of the AFE that allows the overall gain of the AFE to be adjusted. The suggested AFE operates in a wide frequency range of 0.5 Hz to 2 kHz with a high input impedance bigger than 2TΩ, and it is constructed and simulated using a 180nm CMOS process. With the lowest 100-dB CMRR and low input-referred noise of 1.8 µVrms, the AFE can achieve low noise efficiency. EEG signals can be acquired with this AFE system, which is very useful for detecting epilepsy and seizures.

Full Text:

PDF


DOI: http://dx.doi.org/10.21553/rev-jec.339

Copyright (c) 2023 REV Journal on Electronics and Communications


Copyright © 2011-2024
Radio and Electronics Association of Vietnam
All rights reserved